Performance Characteri stics
Bandwidth 25 MHz 40 MHz
Real time sample rate 250 MSa/s 500 MSa/s
(two channels interleaved)
Channels 2
Display 1/4 VGA 1/4 VGA
Monochrome LCD Color LCD
Rise Time <14 ns <8.8 ns
Record Length* 4000 points
Vertical Resolution 8 bits
Vertical Sensitivity 2 mV - 5 V/div
DC gain accuracy ±3.0 %
Maximum Input Voltage 300 Vrms, CAT II (between signal and reference
BNC connector)
Position Range 2 mV – 100 mV range ±2 V
200 mV - 5 V range: ±40 V
Bandwidth Limit - 20 MHz
Time Base range 25 ns/div – 50 s/div 10 ns/div – 50 s/div
Timebase accuracy 100 ppm
Input Coupling AC, DC,GND
Input Impedance 1 MÙ in parallel with13 pf
Vertical and Horizontal Zoom Vertically or horizontally expand or compress a live
or stopped waveform
I/O interface USB device port for connection to PC.
(Requires included EasyScope Software for use)
* The instrument displays 2500 points. 4000 points can be retrieved from internal
memory with the included EasyScope application. This feature is supported for a time base settings
range of 2.5ìs/Div-50ms/Div (scan mode is not active)
Acquisit ion Modes
Sample Display sample data only
Peak Detect
Average Waveform averaged, selectable from
4,16,32,64,128,256
Scan Mode For time base settings 0.1 s/div-50 s/div
Trigger System
Trigger Types Edge, Pulse Width, Video*
Trigger Modes Auto, Normal, Single
Trigger Coupling AC, DC, LF reject, HF reject
Trigger Source CH1, CH2, AC line, Ext, Ext/5
*Support formats PAL/SECAM, NTSC. Triggers on odd or even field, all lines or line
number
Cursors
Types Amplitude, Time
Measurements V, T, 1/ T
Automatic Waveform Measurement
Time Rise time, Fall Time, Cycle Frequency, Period, Positive
Pulse Width, Negative Pulse width
Voltage MAX, MIN, Peak-Peak, Average, Vrms
Frequency Hardware counter provides frequency readout of
trigger source with 6 digit resolution
Waveform Math
Math function FFT, add, subtract, multiply, divide
FFT Windows: Hanning, Hamming, Blackman, Rectangular
1024 sample points
Autoset Single button automatic setup of both channels for
vertical, horizontal and trigger systems
Displ ay
Display Mode 1/4 VGA (5.7”) monochrome LCD (320x240) with
adjustable contrast and inverse video
Display Types Point, Vector
Persistence Off, 1 s, 2 s, 5 s, infinite
Waveform Interpolation Sin(x)/x, Linear
Format YT and XY
Power Requirement s 100-240 VAC, 50 VAmax, 45 Hz to 440 Hz
Envi ronment al
Temperature Operating: 0º C to +55º C
Nonoperating: -40º C to +70º C
Humidity Operating: 95 %RH, 40º C
Nonoperating: 90 %RH, 65º C
Altitude Operating to 4000 m
Pollution Degree Pollution degree 2 for indoor use only.
Elect romagnetic compat ibili ty and Safet y
EMC This oscilloscope is in compliance with council EMC
directive 2004/108/EC
Safety EN61010-1:2001
General
Dimensions 290 mm x 150 mm x 300 mm
Width x Height x Depth 11.4 in x 5.9 in x 11.8 in
Weight 4.6 kg (10 Lbs)
Accessories
Supplied: User Manual, 10:1 Probe set (2 pieces), Power cord , USB interface cable,
EasyScope Software Installation disk
Optional: PR 37A 10:1 Probe, PR 32A Demodulator Probe, PR 55 High Voltage
Probe Þ Þ Þ
3 www.bk prec ision.com
Digital Storage Oscilloscopes
Models 2530 & 2532
Specifications
2530 2532
models
One Year Warranty
|
|