You may also be interested in
Anemometer, High-voltage probe, Power supply, Converter, Digital ph meter
Text version of the page
Accessories
For your nearest distributor, call 800-462-9832 or www.bkprecision.com 51
SUPPLIED: Instruction Manual, Two PR-37A x1/x10/Ref. Probes or equivalent,
AC Power Cord, Spare Fuse
OPTIONAL: PR-32A Demodulator Probe, PR-46A x10 Probe, PR-37A
x1/x10/REF. Probe, PR-100A x100 Probe, PR-55 High Voltage
x1000 Probe, LC-210A Carrying Case
2190B
VERTICAL AMPLIFIERS (CH 1 and CH 2)
Sensitivity 5mV/div to 5 V/div. 1 mV/div to 1V/div (at X5 MAG)
Attenuator 10 calibrated steps in 1-2-5 sequence.
Vernier control provides fully adjustable sensitivity
between steps, adjustment range 1/1 to 1/2.5
Accuracy ±3% (±5% at X5 MAG)
Input Impedance 1MÙ +3%
Input Capacitance 25 pF ±10pF
Frequency Response DC: DC to 100 MHz (-3 dB), 50MHz to 80MHz (-5dB)
X5 MAG DC to 20 MHz (-3 dB)
AC 10Hz to 100 MHz (-3 dB)
Rise Time 3.5 ns (Overshoot <5%)
Signal Delay Time Variable
Square Wave Characteristics Overshoot less than 5%, 10 mV/div range
Other ranges within 5% additional
Maximum Input Voltage 400V (DC + AC peak)
VERTICAL AMPLIFIERS
Operating Modes CH 1, CH 2, Dual, Add
Delay Time Between Channels Within 1 ns between CH 1 and CH 2
Crosstalk 30:1 at 100 kHz
SWEEP SYSTEM
Operating Modes
A A sweep
B Delayed B sweep
B TRIGGERED B sweep triggered after delay
A Time Base
Sweep Mode Auto, normal
Sweep Time: 5s to 20ns/div., 23 steps in 1-2-5 sequence
with variable control
Accuracy ±3%
Hold Off Time Continuously variable. Adjustment range from
normal to 1.5 times the sweep time
B Time Base
Delay Method Continuous delay. Triggered delay
Sweep Time 20ns. to 0.5s/div., 23 steps in 1-2-5 sequence
Accuracy ±3%
Delay Time Start point: 0.5 div to + 0.3 div.
End point: 10 div + 1 div
Delay Jitter Within 1/10,000 of full scale sweep time
TRIGGERING
A Trigger
Source CH 1, CH 2, LINE, EXT
Sensitivity 30Hz to 110MHz
TV-V 20Hz - 30kHz
TV-H 3kHz - 100kHz
Slope + or -
B Trigger The A trigger is also the B trigger
EXTERNAL TRIGGER
Input Impedance 1mÙ, 30pF
Maximum Input Voltage 300V (DC + AC peak)
HORIZONTAL AMPLIFIER
X-Y Mode X Axis = CH 1. Y Axis = CH 2
Sensitivity 5 mV/div to 5 V/div, CH 1 and CH 2
Accuracy ±3% calibrated position, ±6% using x10 MAG
Frequency Response DC to 2 MHz (-3dB)
CH2 (Y) OUTPUT
Output Voltage Approx. 100mV/div open circuit
Approx. 50 mV/div into 50Ù
Freq. Response 50 Hz to 30 MHz.
Output Impedance approx. 50Ù
CRT
Type Rectangular with integral graticule
Display Area 8 x 10 div (1 div = 1 cm)
Accelerating Voltage 15kV
Phosphor P31
Scale Illumination None
Trace Rotation Electrical, front panel adjustable
Z Axis Sensitivity: 3 V or greater, TTL level.
(Intensity Modulation) Negative polarity increases brightness
Input Impedance 15 kÙ
Usable Freq. Range DC to 3.5 MHz
Maximum Input Voltage 20 V (DC + AC peak)
CAL/Probe Compensation
Waveform Positive going squareware
Output Voltage 0.5 V p-p ±3%
Frequency Approx. 1kHz
Duty Cycle 50 ± 5%
Power Requirements 100/120/220/240/ VAC ±10%, 50/60 Hz,
approximately 55 W
Dimensions (HxWxD) 12.76 x 15.68 x 5.2" (324 x 398 x 132 mm)
Weight 18.7 lbs (8.5 kg)
ENVIRONMENT
Within Specified Accuracy 50° to 95°F (10° to 35°C), 85% maximum RH
Full Operation 32° to 104°F (0° to +40°C), 85% maximum RH
Storage -4° to 158°F (-20°to +70°C)
Three Year Warranty
Other Specifications
Specifications model
2190B
¡ Dual time base oscilloscope
(2 channel)
¡ 5mV/division sensitivity
¡ Sweeps to 5ns/division
¡ 23 calibrated ranges, main time base
¡ Signal delay line
¡ 15 kV accelerating voltage
¡ Channel 2 output
on 10mV/Div setting
|
|