Lake MonitorsFlowStat Lake Monitors FlowStat
Perfect monitoring solution forChillers/Cooling Circuits, HVAC,
Medical Equipment, Batching and Industrial process control applications.
® ® –TurbineFlow Sensor –TurbineFlow Sensor
CHOICE OF THREE PORT SIZES SIMPLY PLUMB AND APPLY POWER Select from 1/2" , 3/4" or 1" porting to meet system requirements. Comes factory calibrated to your flow range specifications. EASY MAINTENANCE AND CLEANING VALUE PRICING Has only one moving component, the impeller.Cleaning and maintenance may be performed
without removing the sensor from the piping. Combined with low cost operation and mainte-nance, equals better bottom line savings for your
operation. HERMETICALLY ENCAPSULATED CIRCUITRY Withstands the harshest environments. SEVERAL OUTPUTS AVAILABLE >
ENGINEERING SPECIFICATION The standard interface is a 2-wire, 4-20mA current loop. Sensor signal may be transmitted
on a low cost wire without degradation. Pulse
and 0-5 VDC are also available. CONNECTS DIRECTLY TO YOUR FLOWMONITORING INSTRUMENTS >
THE FLOWSTAT SENSOR SHALL: Can be connected directly to analog acquisitioncards, chart
industrial recorders or other monitoring
instruments, without external signal conditioning. >
? Have only one moving component. ? Be calibrated to user specified flow range. ? Have a measuring accuracy of 2% of full scale. ? Have hermetically encapsulated circuitry. ? Be Lake Monitors Number C_ _ - _ _ - _ _ for
the Current output, Number P_ _ - _ _ - _ _ for
the pulse Output version, and Number V_ _ - _ _ - _ _ for the 0-5 VDC option. www.lakemonitors.com