You may also be interested in
Spectrometer, Correlator, Light source, Impulse generator, Miniature spectrometer
Text version of the page
ASP-150C(ASP-150T) technical specifications
Avesta Project Ltd.
P. N. Lebedev Physical Institute,
KRF-2 building, Troitsk, 142190,
Moscow region, Russia
Tel: +7 495 967-94-73
+7 495 745-22-05
+7 4967 51-00-78
Fax: +7 495 646-04-95
e-mail: fs@avesta.ru
www.avesta.ru
ASP-150C(ASP-150T)
Optical scheme Cherny-Turner
Spectral range for choosing the registration
domain, nm
190-
1100
190-
1100
190-
1100
190-
1100
190-
1100
190-
800
190-
450
190-
300
Grating, grooves/mm 200 300 400 600 1200 1800 1800 II
order
1800 III
order
Registered spectrum region width, nm* 950 640 480 315 145 90 40 25
Spectral resolution, nm 0.6 0.42 0.3 0.2 0.09 0.06 0.025 0.017
Dispersion, nm/mm 33.2 22.4 16.8 11.0 5.1 3.1 1.4 0.9
Input slit, ėm 15
Focal length, mm 150
Focal length of the camera objective, mm 150
Relative aperture 1:13
Array Toshiba 1205D Toshiba 1304AP Hamamatsu
S8378-1024Q
Number of cells 2048 3648 1024
Cell width, ėm 14 8 24
Cell height, ėm 200 200 500
Minimum exposure time, ms 7 7 2,1
Maximum exposure time, s 4 0,5 5
Sensitivity, V/lx*s 80 160 22(4,4)**
Antiblooming*** yes no yes
Meansquare noise of reading, counts of ADC 5,4 3,5 16(4,4)**
ADC 12 bit, 4096
counts
12 bit, 4096
counts
14 bit, 16384
counts
Dynamic range 1000:1 1000:1 1000:1
(4000:1)**
Fiber input****
Synchronization external/internal
PC connection USB
Dimensions, mm 76x115x241.5 (ASP-150T); 76x115x204.5 (ASP-150C)
Weight, kg 1.6
* - The necessary width and center of the registered range should be specified when placing an order. The ASP-150T has an opera-tional
range adjustment function.
** - The HAMAMATSU arrays provide for sensitivity control via special driving signal, that can set either low () or high sensitivity
mode.
*** - Antiblooming - the property of the CCD to prevent the charge flow from the neighboring overexposed pixels.
**** - For models ASP-150CF and ASP-150TF.
FC, 400 ėm UV fiber cable included (1 m.)
|
|