A Aurora
Spectrometer
Optics
Electronic switchable single/double beam, including low UV range. Narrow beam optical design. Aberration corrected 30cm Czerny-Turner monochromator with software controlled wavelength selection and optimization.
Dimensions
W 65 x D 55 x H 37cm
Band Pass
Adjustable continuous bandwidth for 0.1-2nm
Grating
50 mm x 50 mm diffraction grating with 1200 lines/mm
Wavelength Range
190-900 nm controlled by software
Wavelength Scan / Optimization
From the Wavelength Range in any selected range with Zoom Function. Automatic wavelength optimization with graphical display.
Background Correction
Standard Self reversal (Smith-Hieftje) for accurate correction
(Optional: Upgradeable to have D2 or both D2 /Self reversal. Electronic
modulation with deuterium current control and aperture attenuation.)
PMT
Standard PMT from 185-900 nm, automatic gain control.
Drift Correction
ADC (Automatic Drift Correction) permits the full intensity of the light beam to pass through the sample resulting in superior detection power. No HCL preheat is required.
Light Source
5 lamp motorized turret with independent power supply. Automatic alignment and lamp selection.
Flame Atomizer
Gas control
Automatic changeover between different flame systems.
Spray Chamber
ABS spray chamber, inert to acids and organic solvents (Optional: PTFE spray chamber)
Burner
Corrosion resistant Titanium burner heads for both air/acetylene and nitrous-oxide/acetylene.
Gas Requirements
1. Air dry, oil and dust free @ 50-80 psi; Typical flow rate 10.0 12.0 L/min
2. Acetylene (99% pure) @ 8-10 psi; Typical flow rate 0.5 8.0 L/min
3. Nitrous-oxide (99.5 % pure) @ 50-80 psi; Typical flow rate 2.0 10.0 L/min
Flow-rate / Control
Manual (Optional: Programable gas control)
Sample Introduction
Automatic on-line dilution and preparation of calibration curves from a single standard.
Fume Exhaustion
Conventional laboratory extraction system.
Safety System
Safety Interlock mechanism that monitors the burner type, burner fitted, gas pressures. Flame on/off sensors; Power failure automatic shutdown. Pressure relief plug and flame shield to protect from UV radiation.
Optional Flame / VG XYZAuto sampler/Flame/GF Automatic Switch
Auto sampler design
Free movement in XYZ directions, Adapt to any sample container.
Sample Introduction
When combined with on-line dilution the flame autosampler offers, calibration from a single standard and automatic dilution of over-range samples.
F/GF Automatic Switch
Automatic Control by Computer, No Manual assembly requires.
Vapor and Hydride Generator
Absorption Cell
Constant temperature electrothermally heated quartz absorption cell with temperature range of ambient to 1000°C.
Pump
Continuous flow peristaltic pump with speed control.
Accessories
High efficiency mixing section, gas-liquid separator, tubing and connectors included.
Performance
Cold vapor and hydride generation determinations of Hg, As, Se, Te, Bi, Sb, Sn, Cd, Zn, Ge, Pb and other hydride-forming metals at sub-trace levels.
For North American Sales: Aurora Biomed Inc. email: info@aurorabiomed.com website: www.aurorabiomed.com
9001
For International Sales: Aurora Instruments Ltd. email: info@aurora-instr.com website: www.aurora-instr.com
REGISTERED
Address: 1001 E. Pender St., Vancouver BC Canada V6A 1W2 Phone: 1.800.883.2918; 604-215-8700 Fax: 604-215-9700
AI500 Atomic Absorption Spectrometer

pageCatalog pdf di En 2012-05-21-27