Catalogue Ci5000 Weather-Ometer
www.atlas-mts.com
print switch display
Page / 8
Request
a Quote
Atlas Material Testing Technology - 332501, 330803, 89345, 154755, 376128, 158915, 101181
/ 8
See other catalogues for Atlas Material Testing Technology
Text version of the page

STANDARDS & SPECIFICATIONS

Physical Dimensions

Height 198 cm (78 in) Width 160 cm (63 in) Depth 127 cm (50 in)

Standard Features

Floor Space 212 cm (83 in) x 293 cm (115 in) Including Access Area TFT Full Color 10.4" Touch read out Professional control panel Display of All Test Parameters Total Exposure Area 11000 cm
2 ®ÊÊ Direct Setting and Control of Irradiance
®ÊÊ Direct Setting and Control of BPT/BST
®ÊÊ Direct Setting and Control of Relative Humidity

Electrical Specification – Standard Voltage

®ÊÊ Direct Setting and Control of Specimen and Chamber Air Temperature (Dry Industrial bulb) Wiring Connections 3 Phase, 3 Wire w/ Ground (3/PE)
®ÊÊ Display of Diagnostic Messages Operating Voltage Range 440-480 VAC Phase to Phase
®ÊÊ 14 (Standard Voltage) and 12 (Low Voltage) Factory Pre-Programmed Test Methods Maximum Current 60 Amps
®ÊÊ Space for Several Custom Programs Frequency 50/60 Hz
®ÊÊ Multi-Language Capability (English, French, German, Spanish, Japanese, Chinese, Korean)SmartDamper Maximum Power 24 kW

Low Voltage Model

Wiring Connections 3 Phase, 4 Wire w/ Ground (3/N/PE) SmartLight Monitor
Incoming voltage requirements are the same as older Atlas equipment (Ci65 series) with similar sample capacity*Minimizes the need for electrical facilities upgrades Upgradable to standard model in the field if future testing needs change

Weight

Streaming Data Output via SD Card or RS232 or Both Simultaneously (includes SD Professional card reader)Air heating resistor Main Power Protection switch switch Xenon Lamp Cooling System Air Intake Dust filtration Three-tier Specimen Rack Water Purity Indicator Calibrated Xenon Reference Lamp Chamber Viewing Door 316 Grade Stainless Steel Tester chamber Universal Electrical Configurations to Meet Local Frequency, Voltage, and Electrical RequirementsMeets CE, UL, CSA, ISO and EN Compliance Operating Voltage Range 340-415 VAC Phase to Phase Weight of Fully Skidded and Wrapped Ci5000 943 kg (2080 lbs) Maximum Current 63 Amps Weight of Ci5000 without Skid Frequency 50/60 Hz 807 kg (1780 lbs) Maximum Power 24 kW
* Contact your Technical Service representative to review incoming electrical configurations. Minor upgrades may be required

Water Consumption Electrical Specification – Low Voltage

Pressure 138-344 kPa (20-30 psi)

Optional Features

Wiring Connections 3 Phase, 3 Wire w/Ground (3/PE) Flow Rate (max
* ) Deionized Tap Water Water @18.5° CHumidification 6 Channel Printing Chart recorder Dual ASTM/BPT and DIN/BST Black Panel Temperature Measurement/Control Including ASTM/BPT and DIN/BST SensorsMonitoring of Second Wavelength LiquiAir Operating Voltage Range 240 VAC Phase to Phase 0.2 l/min Specimen Spray 0.2 l/min Maximum Current 85 Amps Rack Spray 0.2 l/min Frequency 50/60 Hz Xenon Lamp Cooling @ 4000W
Maximum Power 14 kW 1.5 l/min
* Typical water usage will be less. Tap water requirements for lamp cooling with the LiquiAir will be near zero.

1415