| | Standard connector type | N female | |
| | Optional connector types | 3.5 mm female, 3.5 mm male, GPC-7, N male | |
| | Measurement port characteristics | Connector | Configuration | Frequency (MHz) | Directivity (dB) | Source match (dB) | Load match (dB) | |
| | 3.5 mm (MS4600/11S) (MS4600/11SF) | Ports 1 and 2 MS462xB MS462xD | 10 to 1000 1000 to 3000 3000 to 6000 6000 to 9000 | >46 >44 >38 >37 | >44 >41 >39 >36 | >46 >44 >38 >37 | |
| | Ports 3 and 4 MS462xB/Opt3x MS462xD | 10 to 1000 1000 to 3000 3000 to 6000 6000 to 9000 | >44 >42 >37 >36 | >42 >40 >37 >35 | >44 >42 >37 >36 | |
| | N-Type Standard N(F) (MS4600/11NM) | Ports 1 and 2 MS462xB MS462xD | 10 to 1000 1000 to 3000 3000 to 6000 6000 to 9000 | >46 >44 >38 >37 | >44 >41 >39 >36 | >46 >44 >38 >37 | |
| | Ports 3 and 4 MS462xB/Opt3x MS462xD | 10 to 1000 1000 to 3000 3000 to 6000 6000 to 9000 | >44 >42 >37 >36 | >42 >40 >37 >35 | >44 >42 >37 >36 | |
| | GPC-7 (MS4600/11A) | Ports 1 and 2 MS462xB MS462xD | 10 to 1000 1000 to 3000 3000 to 6000 6000 to 9000 | >46 >44 >38 >37 | >44 >41 >39 >36 | >46 >44 >38 >37 | |
| | Ports 3 and 4 MS462xB/Opt3x MS462xD | 10 to 1000 1000 to 3000 3000 to 6000 6000 to 9000 | >44 >42 >37 >36 | >42 >40 >37 >35 | >44 >42 >37 >36 | |
| | Frequency range | MS4622A/B/D, 10 MHz to 3 GHz MS4623A/B/D, 10 MHz to 6 GHz MS4624A/B/D, 10 MHz to 9 GHz | |
| | Frequency resolution | 1Hz | |
| | Frequency stability (with internal time base) - aging | <5x10-6 / year | |
| | Temperature | <5x10-6over +15°C to +50°C | |
| | Power output range | MS4622A Transmission/Reflection Test Set +10 to -85 dBm MS4622B Active Reversing Test Set +10 to -85 dBm MS4622B (Opt 3) w/ 2nd Source, 3rd Test Port & S/A. +10 to -85 dBm MS4622B (Opt 4) w/ Noise Figure +7 to -85 dBm MS4622B (Opt 6) w/ 3rd Test Port +10 to -85 dBm MS4622D Balanced/Differential 4-Port +10 to -85 dBm | |
| | MS4623A Transmission/Reflection Test Set +10 to -85 dBm MS4623B Active Reversing Test Set +7 to -85 dBm MS4623B (Opt 3) w/ 2nd Source, 3rd Test Port & S/A +7 to -85 dBm MS4623B (Opt 4) w/ Noise Figure (3 GHz only) +5 to -85 dBm MS4623B (Opt 6) w/ 3rd Test Port +7 to -85 dBm MS4623D Balanced/Differential 4-Port +7 to -85 dBm | |
| | MS4624A Transmission/Reflection Test Set +10 to -85 dBm MS4624B Active Reversing Test Set +7 to -85 dBm MS4624B (Opt 3) w/ 2nd Source, 3rd Test Port & S/A +7 to -85 dBm MS4624B (Opt 6) w/ 3rd Test Port +7 to -85 dBm MS4624D Balanced/Differential 4-Port +7 to -85 dBm | |
| | Power control range | > 20 dB. The minimum absolute level for power sweep is -15 dBm while the maximum power output for a unit is +10 dBm. | |
| | Source power level | The source power (dBm) may be set from the front panel menu or via GPIB. Port 1 power level is settable from +10 dBm (on the simpler test sets, ranging to +5 dBm on the most complex) to -15 dBm with 0.01 dB resolution. In addition, the Port 1 (& Port 3) power may be attenuated in 10 dB steps using the internal 70 dB step attenuator. Port 3 step attenuator is not available in D models. Port 1 step attenuator is optional in A models. | |
| | Power level accuracy | ±1 dB to 6 GHz, ±1.5 dB to 9 GHz (no flat power calibration applied; full-band frequency sweep at -15 dBm, 0 dBm, and maximum rated power). | |
| | Level test port power | The power at all sweep frequencies is leveled to within ±1 dB. Only port 1 and port 3 (if installed) can be externally leveled. | |
| | Harmonics and spurious | <-30 dBc at maximum rated power (MS4622x and MS4623x) <-25 dBc at maximum rated power (MS4624x) | |
| | Sweep type | Linear, CW, Marker, or N-Discrete point sweep | |
| | Power sweep range | 20 dB (minimum) | |
| | Source #2 (optional) | Frequency range | 10 MHz to 3 GHz (6 GHz or 9 GHz) | |
| | Frequency resolution | 1Hz | |
| | Power level accuracy | ±1 dB to 6 GHz, ±1.5 dB to 9 GHz (no flat power calibration applied; full-band frequency sweep at -15 dBm, 0 dBm, and maximum rated power). | |
| | Harmonics and spurious | <-30 dBc at maximum rated power (MS4622x and MS4623x) <-25 dBc at maximum rated power (MS4624x) | |
| | Sweep type | Linear, CW, Marker, or N-Discrete point sweep | |
| | Power sweep range | 20 dB (minimum) | |
| | | | | | | | | | |