Analyzers catalogue - Anritsu - #485

/ 655


catalogue search
P. 451
P. 452
P. 453
P. 454
P. 455
P. 456
P. 457
P. 458
P. 459
P. 460
P. 461
P. 462
P. 463
P. 464
P. 465
P. 466
P. 467
P. 468
P. 469
P. 470
P. 471
P. 472
P. 473
P. 474
P. 475
P. 476
P. 477
P. 478
P. 479
P. 480
P. 481
P. 482
P. 483
P. 484
P. 485
P. 486
P. 487
P. 488
P. 489
P. 490
P. 491
P. 492
P. 493
P. 494
P. 495
P. 496
P. 497
P. 498
P. 499
P. 500


See other catalogues for Anritsu

Text version of the page

SPECTRUMANALYZERS

Setting rangeLog scale: –100 to +30 dBm; Linear scale: 224 ΅V to 7.07 VUnitLog scale: dBm, dB΅V, dBmV, V, dB΅Vemf, W, dB΅V/m Linear scale: VReference level accuracy:Reference level±0.4 dB (–49.9 to 0 dBm), ±0.75 dB (–69.9 to –50 dBm, 0.1 to +30 dBm), ±1.5 dB (–80 to –70 dBm)

βˆ—

After calibration, at 100 MHz, span: 1 MHz (when RF ATT, RBW, VBW, and sweep time set to AUTO)RBW switching uncertainty: ±0.3 dB (1 kHz to 1 MHz), ±0.4 dB (3 MHz)

βˆ—

After calibration, referenced to RBW: 3 kHzInput attenuator (RF ATT)Setting range: 0 to 70 dB (10 dB steps)

βˆ—

Manually settable, or automatically settable according to reference levelSwitching uncertainty:±0.3 dB (0 to 50 dB), ±1.0 dB (0 to 70 dB)

βˆ—

After calibration, frequency: 100 MHz, referenced to RF ATT: 10 dB ±0.5 dB (100 kHz to 3 GHz, referenced to 100 MHz, RF ATT: 10 dB, +18 to +28°C)Frequency response±1.5 dB (9 to 100 kHz, referenced to 100 MHz, RF ATT: 10 dB, +18 to +28°C)±1.0 dB (100 kHz to 3 GHz, referenced to 100 MHz, RF ATT: 10 to 50 dB) Scale (10 div)Log scale: 10, 5, 2, 1 dB/divLinear scale: 10, 5, 2, 1%/divLinearity (after calibration)Waveform displayLog scale: ±0.4 dB (0 to –20 dB), ±1.0 dB (0 to –70 dB), ±1.5 dB (0 to –85 dB), ±2.5 dB (0 to –90 dB)Linear scale: ±4% (compared to reference level)Marker level resolutionLog scale: 0.01 dB; Linear scale: 0.02% of reference level 2nd harmonic distortion: ≀ –60 dBc (10 to 200 MHz), ≀ –75 dBc (0.2 to 1.5 GHz), ≀ –80 dBc (0.8 to 1 GHz)

βˆ—

Mixer input: –30 dBmSpurious responseTwo signals 3rd order intermodulation distortion: ≀ –70 dBc (10 to 100 MHz), ≀ –80 dBc (0.1 to 3 GHz)

βˆ—

Frequency difference of two signals: β‰₯ 50 kHz, mixer input: –30 dBm 1 dB gain compression β‰₯ –5 dBm ( β‰₯ 100 MHz, at mixer input level) 1 dB gain compression level to average noise level:>110 dB (0.1 to 1 GHz), >110 dB – f [GHz] dB (>1 GHz), >109 dB (0.1 to 1 GHz, at Option 08 pre-amplifier installed),>109 dB – 1.5f [GHz] (>1 GHz, at Option 08 pre-amplifier installed)Maximum dynamic rangeDistortion characteristics (RBW: 1 kHz)2nd harmonic:>72.5 dB (10 to 200 MHz), >80 dB (200 to 500 MHz), >80 – f [GHz] dB (0.5 to 1.5 GHz), >82.5– f [GHz] dB (0.8 to 1 GHz)3rd order intermodulation: >80 dB (10 to 100 MHz), >83.3 dB (0.1 to 1 GHz), >83.3 – (2/3)f [GHz] dB (1 to 3 GHz) Sweep timeSetting range : 20 ms to 1000 s (Manually settable, or automatically settable according to span, RBW and VBW)Accuracy: ±15% (20 ms to 100 s), ±45% (110 to 1000 s), ±1% (time domain sweep: digital zero span mode) Sweep modeContinuous, single Time domain sweep modeAnalog zero span, digital zero span Zone sweepSweeps only in frequency range indicated by zone marker Tracking sweepSweeps while tracing peak points within zone marker (zone sweep also possible) Number of data points501 NORMAL: Simultaneously displays max. and min. points between sample points POS PEAK: Displays max. point between sample pointsDetection modeNEG PEAK: Displays min. point between sample pointsSAMPLE: Displays momentary value at sample pointsDetection mode switching uncertainty: ±0.5 dB (at reference level) DisplayColor TFT-LCD, Size: 5.5 inch, Number of colors: 17 (RGB, each 64-scale settable); Intensity adjustment: 5 steps settable Trace A: Displays frequency spectrumTrace B: Displays frequency spectrum Trace Time: Displays time domain waveform at center frequencyTrace A/B:Displays Trace A and Trace B simultaneously. Simultaneous sweep of same frequency, alternate sweep of Display functionsindependent frequenciesTrace A/BG:Displays frequency region to be observed (background) and object band (foreground) selected from background with zone marker simultaneously at alternate sweep Trace A/Time: Displays frequency spectrum, and time domain waveform at center frequency simultaneously at alternate sweep Trace move/calculation: A β†’ B, B β†’ A, A ↔ B, A + B β†’ A, A – B β†’ A, A – B + DL β†’ A Storage functionsNORMAL, VIEW, MAX HOLD, MIN HOLD, AVERAGE, CUMULATIVE, OVER WRITE Demodulation range: 2, 5, 10, 20, 50, 100, 200 kHz/divMarker display accuracy: ±5% of full scale (referenced to center frequency, DC-coupled, RBW: 3 MHz, VBW: 1 Hz, CW)FM demodulation waveformDemodulation frequency response:display functionDC (50 Hz at AC-coupled) to 100 kHz
FunctionsSweepAmplitude

βˆ—

Range: ≀ 20 kHz/div, VBW: off, at 3 dB bandwidthDC (50 Hz at AC-coupled) to 500 kHz

βˆ—

Range: β‰₯ 50 kHz/div, VBW: off, at 3 dB bandwidth

βˆ—

RBW: β‰₯ 1 kHz usable Input connectorN-J, 50 Ω IF OUTPUT: 10.69 MHz, BNC connectorVIDEO OUTPUT (Y):0 to 0.5 V ±0.1 V (100 MHz, from lower edge to upper edge at 10 dB/div or 10%/div, 75 Ω terminated, Auxiliary signal input andBNC connector)outputCOMPOSITE OUTPUT: For NTSC, 1 Vp-p (75 Ω terminated), BNC connectorEXT REF INPUT: 10 MHz ±10 Hz, β‰₯ 0 dBm (50 Ω terminated), BNC connector Signal searchAUTO TUNE, PEAK β†’ CF, PEAK β†’ REF, SCROLL Zone markerNORMAL, DELTA Marker β†’ MARKER β†’ CF, MARKER β†’ REF, MARKER β†’ CF STEP SIZE, βˆ† MARKER β†’ SPAN, ZONE β†’ SPAN Peak searchPEAK, NEXT PEAK, NEXT RIGHT PEAK, NEXT LEFT PEAK, MIN DIP, NEXT DIP Continued on next page

http://www.anritsu.com 482


pageCatalog pdf di En 2012-02-07-13