OPTICAL MEASURING INSTRUMENTS
/inritsu
• Optical sensors
Model
MA9421A
MA9422A
MA9423A
MA9621A
MA9622A*1
Wavelength range
0.38 to 1.15 urn
0.75 to 1.7 um
1.2 to 1.7 um
Element
SI diode
InGaAs diode
Active area diameter
09.5 mm
09 mm
09.5 mm
01 mm
-
Input
Direct
FC, ST, DIN, HMS-10/A, SC type connector adapter*2
FC, SC, ST, DIN,
HMS-10/A, replaceable connector, PC polishing
Measurement
CW (dBm)
-60 to +20 (0.85 um)
-50 to +20 (0.85 um)
-70 to +10 (0.85 um)
-70 to +3 (1.3 um)
-50 to +23 (1.3/1.55 um)
range
MOD
(dBm)
-65 to +17 (0.85 um)
-50 to +17 (0.85 um)
-75 to +7 (0.85 um)
-75 to 0 (1.3 um)
-55 to +20 (1.3/1.55 um)
Measurement accuracy*3
±5%*4
±5%*5
±5%*6
±5%*7
Temperature range
0 to +50"C (use), —40 to +70"C (storage)
Dimensions and mass
30 (W) x 30 (H) x 37 (D) mm, <100 g
15 (W) x 16 (H) x 140 (D) mm, <200 g
30 (W) x 30 (H) x 37 (D) mm, <100 g
*1: Applicable connector: SM fiber (ITU-T G.652)
Return loss: >40 dB (1.55 ±0.2 um, only when return loss of optical connector: >45 dB)
Polarization dependency: <0.1 dB (1.55 ±0.02 um) *2: Specify one connector among those shown in the specification table.
When no connector and manufacturer's name are specified, FC-type will be mounted and supplied.
Other than the connectors indicated in the table are dealt in special connectors of custom-made. The ordering method of optical connectors are indicated in
the table on page ??. *3: Used with FC-type connectors *4: At -10 dBm, 0.633/0.78/0.85 um CW light mode *5: At -10 dBm, 0.66/0.78/0.85 um CW light mode *6: At -10 dBm, 0.85/1.3/1.55 um CW light mode *7: At -10 dBm, 1.3/1.55 um CW light mode
• MS0907A Return Loss Measurement Unit*1
Applicable fiber
SM (10/125 um, NA 0.1)
Wavelength
1.31 ±0.03 um (25"C)
Measurement range
0 to 40 dB (relative to total internal reflection cord, including output connector reflection)
Measured data display range
0 to 60 dB (relative to total internal reflection cord, excluding output connector reflection)
Measurement accuracy
±1 dB (relative to the reflection, constant temperature)
Optical output connector*2
FC, ST, DIN, HMS-10/A, SC: PC-type
Temperature range
0 to +50"C (use), -40 to +70"C (storage)
Dimensions and mass
90 (W) x 93 (H) x 36 (D) mm, <300 g
*1: Laser Product Safety Standards: IEC 60825-1 Class-1, FDA 21CFR Class-1 *2: Specify one connector among those shown in the specification table.
When no connector and manufacturer's name are specified, FC-type will be mounted and supplied.
Other than the connectors indicated in the table are dealt in special connectors of custom-made. The ordering method of optical connectors are indicated in the table on page ??.
Safety measures for laser products
The MS0908A complies with the optical safety standards in Class 1M of the IEC 60825-1 and the FDA (21CFR 1040.10, USA); the fol­lowing descriptive labels are affixed to the product (FDA label is only affixed to product for export to the USA).
FDA (CFR)
IEC
The maximum output Is Indicated under *1, and the wavelength under *2. Caution: Do not look directly Into the laser beam.