LIFEGARD II with ICG & EtCO2 Data Sheet - ANALOGIC - #1

/ 2


catalogue search
LIFEGARD II with ICG & EtCO2 Data Sheet - ANALOGIC
P. 01
LIFEGARD II with ICG & EtCO2 Data Sheet - ANALOGIC
P. 02
Pages:
LIFEGARD II with ICG & EtCO2 Data Sheet - ANALOGIC


See other catalogues for ANALOGIC

Text version of the page
IFEGARD II I CG- I
The Most Powerful Non-Invasive Monitor in its Class
The Analogic UFEGARD® II is the most pow­erful, non-invasive monitor in its class avail­able today This versatile monitor measures seven vital signs plus up to 17 hemodynamic values and can be used anywhere in the hospi­tal or office environment
• Up To 7 Non-invasive Measurements -ECG (HR), RESP, SpO2, NIBP, TEMP and ICG, orCC>2
• Continuous Non-Invasive Cardiac Output -ICG (Impedance Cardiography) to calculate 17 hemodynamic values including cardiac output
• Microstream® continuous airway CO2 and breath rate
• Up To 25 Measured and Derived Vital Sign Parameters Facilitate a More Comprehensive Patient Assessment
• 3-hour Typical Battery Life for Prolonged Transport Monitoring
• User Interface that Enhances Simplicity of Use, Data Presentation, and Confirms Signal Integrity
• Large Color Display - Exceptional Data Visualization
• Network Ready for Connection to LIFEGARD Vue Central Alarm Annunciation System
• 3- or 5-Lead ECG with Support for 2-Lead Simultaneous Display when Using 5 Leads
r-
CO
Specifications
ECG Performance
Leads: 3 or 5 leads, user selectable Heart Rate Range: 20 to 250 bpm Heart Rate Accuracy: ±1 bpm or ±1% Bandwidth-Normal Monitoring: 0.5 Hz to 40 Hz Bandwidth-Extended Low Frequency Range (user selectable): 0.05 Hz to 40 Hz Display Sweep Speeds: 12.5,25,50 mm/sec ECG Size (Sensitivity): 0.5,1,2,4 mV/cm Lead-Off Detection with display indicator Pacemaker Detection Mode: Indicator on waveform display, user selectable Standards: ANSI/AAM EC13 Differential Input Impedance: > 5 MQ Common Mode Rejection Ratio:
>90dBat50or60Hz Input Dynamic Range: ± 5 mV AC, ± 300 mV DC Defibrillation Artifact Recovery Time:
< 5 sec conforming to AAMIEC13
Respiration Performance
Range: 3 to 150 breaths per minute
Accuracy: ± 3 breaths per minute
Display Sweep Speeds: 6.25,12.5,25 mm/sec
NIBP Performance
Blood Pressure Accuracy: Meets performance standards
of ANSI/AAM SP10(1) Technique: Oscillometric Pulse Rate Range: 40 to 200 bpm Maximum Cuff Pressure: Safety Limited to 300 mmHg Pulse Rate Accuracy: ±2 bpm or ±2% Initial Cuff Inflation is User Adjustable; Factory Default:
160 mmHg (adult mode); 120 mmHg (pediatric mode) Subsequent CuffInflation; Previous; Systolic ±30 mmHg
Blood Pressure Measurement Range
Systolic: 40 to 250 mmHg Diastolic: 20 to 200 mmHg Mean Arterial Pressure: 27 to 220 mmHg Measurement Modes
Auto: Automatic measurements at intervals of 1, 3,5, 10, 30, 60 and 90 minutes Manual: Single measurement initiated by Start/Stop button
STAT: Series of consecutive measurements for 5 minutes
SpO2 Performance
Nellcor® OXIMAX™ Technology %SaturationRange: 1%to100% Pulse Rate Range: 30 to 250 bpm SpO2 Accuracy: 70% to 100% ± 3 digits
with DS100A sensor Pulse Rate Accuracy: ± 3 bpm Temperature Performance Range: 15°C to 45°C (59°F to 113°F) Accuracy: ± 0.1°C (0.18°F) Compatible with YSI 401 and 409 Temperature Probes

pageCatalog pdf di En 2012-02-07-15