Interfaces
RS-232 digital output from the serial port for one of the following: nurse call
transferring data to an external computer transferring patient data and waveforms to LIFEGARD Vue Central System
Physical
Weight: 5.7 kg (125 lb.) excluding
accessories, options, cables Size: 34.3 cm x 25.4 cm x 17.8 cm
(13.5"x10.0"x7.0")
Monitor Environmental
Operating Temperature: 5°C to 45°C (41°F to 113°F) Storage Temperature: - 20°C to 60°C (- 4°F to 140°F) Operating/Storage Humidity: 20% to 90% relative humidity, non-condensing Operating Altitude: 0 to 3,048m (0 to 10,000 ft.) Mechanical Shock: 50G
Power
AC 100 to 240 VAC, 50/60 Hz, 1A Monitor Performance
Display TFT color-active matrix
Screen Size: 210.8 mm x 157.5 mm (83" x 6.2")
264.2 mm (10.4") diagonally Resolution: 640 x 480 pixel View Angles: 70° left / right, 40° up, 70° down Backlight Life: 25,000 hours Waveforms
Up to 4 traces (selectable in any combination of 2 ECG, etCO2,SpO2 plethysmograph, or respiration) Indicators Categorized alarms
(3 priority levels with different tone pitch) Visual alarm notification Heart rate tone Battery status External power LED
Compliance and Approvals
CE marking: Medical Device Directive, 93/42/EEC Safety Standards: IEC 60601-1, IEC 60601-1-1,
UL 60601-1, CAN/CSAC22.2 No 601.1 (M90),
EN 60601-1
Water Resistance: IEC 60529 Classification IPX1 Mechanical Shock: IEC 60068-2-27 Environmental Humidity: IEC 60068-2-78 Environmental Temperature: IEC 60068-2-1,
IEC 60068-2-2
Electromagnetic Compatibility: IEC 60601-1-2
UL60601-1, CAN/CSA-222 No. 601.1 (M90)
IEC/EN 60601-14
IEC/EN 60601-1-8
IEC/EN 60601-2-27
IEC/EN 60601-2-30
IEC/EN 60601-249
EN864 EN865
EN 1060-1 EN 1060-3 ANSI/AAMISP10 ISO 9918
CO2
Complies with EN 864/ISO 9918
Range: 0to99mmHg(0to13kPa), or 13%CO2
whichever is lower. Flow Rate: 50 ±7.5 ml/min
Warm-up Time: 180 seconds max.; typically 30 seconds Gas Sampling Delay Time: 2.7 seconds Resolution
Numeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
Rise Time: 190 ms for adult mode (measured with sample line for humidified ventilation and airway adapter for adult)
Calibration Interval: 4,000 operating hours
Auto Zero Intervals: < 2% duty cycle. Increases if the temperature changes by 8°C from the previous Auto­Zero interval, or the pressure changes by 45 mmHg.
Leak Tightness: < 40 mbar/min with 30% vacuum
Accuracy
0 - 20 minutes: 0 to 38 mmHg ±4 mmHg; 38 to 99
mmHg ±12% of indicated value > 20 minutes: 0 to 38 mmHg ±2 mmHg 38 to 99 mmHg ±5% of indicated value +
0
U.UÔ7C
above 40 mmH
1.0 mmHg
Airway Respiration
I (awRR)
Specifications subject to change without notice.
Range: 0 to 150 breaths/min Accuracy: 0 to 41 breaths/min ±1 breaths/min; 41 to 70
breaths/min ±2 breaths/min; 71 to 100 breaths/min ± 3
breaths/min; > 100 breaths/min ±5% of reading Graphical and Tabular Trends Memory Storage: 72 hours Data Interval: 20 seconds Display Range: 2 hours, 72 hours scrollable Tabular Format: One table for all variables;
six fields per row (time and 5 vital signs) Display Interval: Per NIBP measurement or
15 minutes if no NIBP, or during an alarm
condition
Adjustable vertical scaling when using graphical trends
Stored Data: Trend data for HR, SpO2,NIBP,awRR,
etCO2, and Temp. Any parameter may be graphically trended
DdllCIV ISLUlUdlUI
Internal battery: sealed lead-acid Battery status indicator Operating Time: 3 hours typ. (fully charged battery) Thermal Printer: Speeds: 12.5,25, and 50 mm/sec Weight: 0.4 kg (0.9 lb.) Paper Width: 50 mm
(1) Systolic and Diastolic blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stetho­scope auscultation method, and mean pressure meas­urements determined with this device are equivalent to those obtained by an intra-arterial blood pressure measurement device, within the limits prescribed by the American National Standards Institute and Association for the Advancement of Medical Instrumentation (Electronic or Automated Sphygmomanometers).
LIFEGARD is a registered trademark of Analogic Corporation. Printed in U.S.A.©2005,2007
16-00208-21 Rev 02
Refer to the LIFEGARD II Ordering Guide for additional information.