| | | Performance Data | | |
| | | |
| | | Figure 2 Sensitivity Temperature Dependence | | |
| | | |
| | | | | | | | | | | | I-129- ■un | | | | | IIU | | | | | 00 | | | | | | | | | | ^^^^^ 80 | | -♦-Mean "+05%, conf." | | | | | --1-1-1-1-69- | | "-05% conf." | | | | | | | | | | | | |
| | | Figure 2 shows the variation in sensitivity caused by changes in temperature. Repeatable temperature dependence at elevated temperatures allows more accurate temperature compensation. This data is taken from a typical batch of sensors. The mean and ± 95% confidence intervals are shown. | | |
| | | o o o CM o <D O m | | |
| | | |
| | | 0s | | |
| | | |
| | | |
| | | |
| | | -30 | | -20 | | -10 | | 0 | | 10 | | 20 | | 30 | | 40 | | 50 | | |
| | | |
| | | Temperature (oC) | | |
| | | |
| | | Figure 3 Zero Temperature Dependence | | |
| | | |
| | | -5 | | |
| | | Figure 3 shows the variation in zero output caused by changes in temperature, expressed as ppm gas equivalent. This data is taken from a typical batch of sensors. The mean and ± 95% confidence intervals are shown. | | |
| | | |
| | | m) p p 0) TO > (e ut pt ut O a er N | | |
| | | 50 | | |
| | | |
| | | |
| | | |
| | | —4 | | |
| | | |
| | | Temperature (oC) | | |
| | | |
| | | Figure 4 Overgas Performance | | |
| | | |
| | | 200000 180000 160000 140000 120000 100000 80000 60000 40000 20000 0 | | |
| | | Figure 4 shows sensor output for increasing concentrations of CO, up to twice the specified overgas range. Data are derived from eight sensors taken from a typical production batch. This stepped overgas test shows the robustness of the sensor with fast response and straight plateaus at each step. | | |
| | | 4,000 ppm | | |
| | | |
| | | 3,000 ppm | | |
| | | |
| | | 2,500 ppm | | |
| | | ut tp ut O | |
| | | |
| | | 2,000 ppm | | |
| | | |
| | | 1,600 ppm | | |
| | | |
| | | |
| | | 1,200 ppm | | |
| | | |
| | | 1,000 ppm | | |
| | | |
| | | 400 ppm | | |
| | | |
| | | |
| | | |
| | | 0 | | 2000 | | 4000 | | 6000 8000 10000 Time (sec) | | 12000 | | 14000 | | 16000 | | |
| | | | |
| | | |
| | | For further information on the performance of this sensor, on other sensors in the range or any other subject, please contact Alphasense Ltd. or visit our web site at "www.alphasense.com". | | |
| | | |
| | | In the interest of continued product improvement, we reserve the right to change design features and specifications without prior notification. The data contained in this document is for guidance only. Alphasense Ltd accepts no liability for any consequential losses, injury or damage resulting from the use of this document or the information contained within it (©ALPHASENSE LTD) Doc. Ref. TDS/COD4/lssue 122 | | |
| | | |
| | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |