Figure 2 shows the variation in sensitivity caused by changes in temperature.
This data is taken from a typical batch of sensors. The mean and ± 95% confidence intervals are shown.
o o
o
CM O
<D O
m
-I—I
H—I
CO
0s
-10
0
10
20
30
50
-30
-20
40
Temperature (oC)
Figure 3 Zero Temperature Dependence
Figure 3 shows the variation in the zero by changes in tempera­ture.
This data is taken from a typical batch of sensors. The mean and ± 95% confidence intervals are shown.
£= Q
re
4— —
e r
-)—t
u p
t
u
o o
er
N
Temperature (oC)
Figure 4 Response to Exposure to 2% CO
1600000
Figure 4 shows the excellent response to step changes in CO concentrations from zero to 2% CO by volume.
This data is taken from a typical batch of sensors.
2% CO
1400000
1200000
1000000
t
u p
t
u
o
800000
600000
400000
200000
0
6000 8000 Time (seconds)
0
2000
4000
10000
12000
14000
16000
For further information on the performance of this sensor, on other sensors in the range or any other subject, please contact Alphasense Ltd. For Application Notes visit "www.alphasense.com"
In the interest of continued product improvement, we reserve the right to change design features and specifications without prior notification. The data contained in this document is for guidance only. Alphasense Ltd accepts no liability for any consequential losses, injury or damage resulting from the use of this document or the information contained within it (©ALPHASENSE LTD) Doc. Ref. TDS/COAF/ Issue 12