specifications
PERFORMANCE
Measurement Ranges (parts per million) Single-range Three-range
Outputs
4-20 mADC and 0-2 VDC standard. RS-232C or RS-485 optional. TTL output with 3 range instruments.
Three (3) SPDT Form C contacts rated 10 A @ 30 VDC/115/230 VAC. User selectable for fail-safe or non fail-safe.
0-10,000 0- 5,000 0-1,000 0-100 0- 50
0-10
Accuracy1 Linearity Response Time
Sensor Type
0-100/1,000/10,000 0- 50/500/5,000
0-10/100/1,000 0- 5/50/500
0-1/10/100
Oxygen Alarm
Relays
Instrument Status Rated identically to the O2 relays. Alarm
Audible Alarm Internal buzzer with audible cancel.
SAMPLE GAS CHARACTERISTICS Flow Rate 1.0 to 2.0 SCFH (0.5 to 1.0 liter/min).
Sample Gas Temp. 32° to 122° F (0° to 50° C).
±1% of full scale (±5% fs 0-1ppm range) ±1%
90% of full scale response in less than 10 seconds (typical). Recovery time for ranges <10 ppm will depend on the prevailing conditions.
Patented Ambient Temperature Electrochemical Sensor.
Standard.
Sample Gas Pressure
0.1 to 1.0 psig (1.4 to 14.2 kg/cm2).
Entrained Solids <3 mg/ft3: no in-line filter required.
>3 mg/ft3: in-line filter is required.
Hydrocarbon Mist <0.7 mg/ft3: no in-line filter required.
>0.7 mg/ft3: in-line filter is required.
CONSTRUCTION2 Electronics Control Polycarbonate with a hinged clear front cover, Unit (Benchtop) rated NEMA 4X (IP 66).
Control Unit 9.45 in. (240.5 mm) length.
Dimensions 6.50 in. (165.1 mm) width.
6.20 in. (157.5 mm) depth.
Gas Connections 1/4" compression fittings.
Sensor Mounting Local or optional remote mounting.
Weight 11lbs (4.98 kg) (Benchtop Config.)
1 Stated at constant temperature and constant pressure.
2 Other mechanical configurations available-consult factory.
Temp.
Compensation
Operating Temperature
32° to 122° F (0° to 50° C).
<32° F use heated sensor enclosure.
>122° F cooling of sensor required.
3 years electronics,1 year sensor.
Warranty
ELECTRICAL Display
Resolution
0.4" (10.2 mm) high, 4-1/2 digit LCD.
0.1 ppm for ranges <100 ppm. 1.0 ppm for ranges >100 ppm.
Input Power 115/230 VAC, 50-60Hz or 18-32 VDC.
NICAD battery backup available as an option for some models.
Alpha Omega Instruments reserves the right to change or modify its product specifications without notice.
Represented by:
Fax: 401-333-5550 Email: info@aoi-corp.com
Visit our Web Site: http://www.aoi-corp.com
30 Martin Street Cumberland, RI, USA 02864
Telephone: 800-262-5977 401-333-8580
ALPHA OMEGA INSTRUMENTS
9/97