LC PAL - Versatile LC injection system for modem chromatography labs - Alpha MOS - #2 |
|
You may also be interested in
Sampler, Chromatograph, Thermal desorber, Gas chromatograph, Electronic nose
Text version of the page
injection procedure of a LC chromatographer with the precision and
throughput of a robotic liquid handling system. Your valuable samples are
aspirated with a conventional syringe and are injected directly into the
sample loop. No error prone teflon tubing or transfer lines are involved
during the sample loading process.
Advanced performance
The LC PAL provides outstanding performance and maximum flexibility
in a LC sample processing system. The standard unit provides sample
capacity for 216 2ml vials or 4 standard- or deepwell microplates in any
combination. An injection volume range of 100nl up to 2500ěl meets the
requirements of virtually any application. The fast cycle time of 20-30
seconds from injection to injection keep pace with the demands of fast
LC and LCMS Flow Injection Analysis (FIA). Additional flexibility can be
added by using the optional 4- or 10-port valves. This enables complex
column switching procedures or internal loop configurations.
Intelligent system control
CTC Analytics' user interface is consistent throughout the whole product
range. Every instrument type operates with the same easy-to-learn, easy-touse
handheld keyboard. This makes using CTC‘s injection systems fast and
efficient. It‘s simplicity of operation translates directly into improved
productivity for laboratories. For individual application requirements the PC
based Windows 95/98/NT control software Cycle Composer is available.
Specifications LC PAL
System Type
XYZ robot with syringe only concept, no tubing in sample path
Syringe sizes
10, 25, 100, 250, 1000, 2500 ěl
LC Injector
Electrically actuated fast switching 6 port valve
Sample Capacity
1 Trayholder for up to:
4 Standard Microplates (96/384 wells )
or 4 Deepwell Plates
or 216 2ml vials
Needle, Syringe, Valve Cleaning
Wash Station for 2 different solvents
(optional Fast wash station, 2 different solvents)
Local Control
Control panel with 4 function keys, Graphical LCD Display
Unique scroll knob for teach functions
Remote Control
Windows 95/98/NT® based PC software
Electrical Control
2 RS 232C ports
3 TTL Input / 1 Opto Coupler Input
2 Relay Output
Dimensions
L: 534mm D: 385mm H: 575mm
Weight
~ 8kg (without accessories)
Power Requirements
97.5-264 VAC, 45-66Hz
Specifications are subject to change without notice
Flexible utilization of microplates with 2ml vials on the same
Trayholder
Direct injection into a sample loop
Optional Fast Wash Station for high throughput applications
LC PAL Options
Thermostatted Sample Trays
4 - and 10 - port electrically actuated fast switching valve
Distributed by:
Where design meets performance
CTC Analytics AG
Industriestrasse 20
CH-4222 Zwingen
Switzerland
Tel: +4 1 61 7658100
Fax: +4 1 61 7658199
E-mail: info@ctc.ch
|
|