The Next DECADE Ahead!
DECADE II
LC-EC Workstation
DECADE II
LC-EC Workstation
Antec Leyden bv
Industrieweg 12,
2382 NV Zoeterwoude,
The Netherlands
Tel: + 31 71 581 3333
Fax: + 31 71 581 3334
mailbox@antecleyden.com
www.antecleyden.com
Ranges 10 pA – 200 ìA in 1, 2, 5 steps
Filter (cut off) Advanced Digital Filter, 0.5 - 0.001 Hz, 1, 2, 5 steps
Noise better than 2 pA with a dummy cell (load of 300 MÙ and 0.5 ìF),
filter off, Ec +800mV and temperature of 30°C.
Range 10 nA – 200 ìA in 1, 2, 5 steps
Filter (cut off) Advanced Digital Filter, 0.5 - 0.001 Hz, 1, 2, 5 steps
Pulse times t1: 100 - 2000 ms; t2: 0 - 2000 ms; t3: 0 - 2000 ms in 10 ms steps
Sample times 20, 40, 60, 80 and 100 ms
Range 10 nA – 200 ìA in 1, 2, 5 steps
Scan rate 1 - 50 mV/s in 1, 2, 5 steps
Cycle half, full or continuous
DC mode (5 files) and pulse mode (4 files), cycle time, number of cycles and oven
temperature. Time-based control of 50 time points as to range, filter, output contacts
(2 TTL, 2 relays), auto zero, board id, offset, valve position (if present) and cell potential.
Mains, Output, 2 Connectors 15 pins (A, B), manual valve (C), RS232C connector
Dimensions 44 (D) x 22 (W) x 44 (H) cm = 17.3” (D) x 8.7” (W) x 17.3” (H)
Weight 14.6 kg (32 lbs) without flow cell and column
Power 100-240 VAC, 50/60 Hz, 260 VA max., auto-sensing
Operating modes DC, PAD and Scan
Potential range between +2.00 and -2.00 V in 10 mV increments
Output between +1 and -1 V or between +10 and -10 V
(20 bit D/A converter)
Offset between +50% and -50% of max. output voltage, 5% steps
Event marker pulse of 10% of max. output
Auto zero triggered by keyboard, rear panel TTL, or RS232C control
RS232C full parametric instrument control, data acquisition at 1, 2, 5
and 10 Hz
Injector sensor starts system clock at injection
Oven height 37 cm, from 7°C above ambient to 45°C, accuracy 0.5°C,
stability 0.1°C; accommodates column and flow cell(s)
Diagnostics LCD screen, keyboard and noise (internal dummy cell)
Service mode system settings and calibration parameters
Configuration menu for system customisation and optimisation
Firmware upgradeable via PC (RS232), flash technology
Environmental operating temperature: 4 – 40°C, RH 20 to 80%,
non-condensing
Physical specifications
Specifications DECADE II
General
specifications
DC mode
PULSE mode
SCAN mode
AUTO mode
Rear panel
I/O connections
|
|