Agilent 2-Port and 4-Port PNA-L Network Analyzer
40Pages

{{requestButtons}}

Catalog excerpts

Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 1

PNA-L Network Analyzer Technical Specifications Agilent Technologies

Open the catalog to page 1
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 2

Documentation Warranty THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS...

Open the catalog to page 2
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 5

This is a complete list of the technical specifications for the N5239A, N5231A and N5232A PNA-L Series network analyzers with the following options: 2-Port, All Models Option 200 - 2-port base model with standard test set. Option 216 - To base model, adds front-panel jumpers, and source attenuators (extended power range). 4-Port N5231A or N5232A Option 400 - 4-port base model with standard test set. Option 416 - To base model, adds front-panel jumpers, and source attenuators (extended power range). See block diagrams for all models and options beginning on page 37. Definitions All...

Open the catalog to page 5
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 6

Corrected System Performance The specifications in this section apply for measurements made with the N5231A and N5232A PNA network analyzers with the following conditions: ■ No averaging applied to data ■ Isolation calibration with an averaging factor of 8 System Dynamic Range and Receiver Dynamic Range ■ System Dynamic Range is defined as the specified source maximum output power (spec) minus the noise floor ■ Extended Dynamic Range at Direct Access Input is defined as the specified source maximum output power (spec) minus the direct receiver access input noise floor (spec). ■ Receiver...

Open the catalog to page 6
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 7

Table 1a. System Dynamic Range (dB), All Models Description Test Port Test Port Direct Receiver Access Port Test Port Test Port Direct Receiver Access Port Table 1b. System Dynamic Range (dB), N5231A or N5232A Description Test Port Test Port Direct Receiver Access Port Test Port Test Port Direct Receiver Access Port

Open the catalog to page 7
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 8

Table 2. Receiver Dynamic Range (dB), N5239A, N5231A or N5232A - Typical Description Options 200, 216 Options 400,416 N5239A, N5231A and N5232A Corrected System Performance, All Options Note: For any Sii reflection measurement: Applies to the N5239A/1A/2A Option 200, 216,400,416 analyzers, 85131F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition: Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Open the catalog to page 8
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 9

Source Match Load Match 1 Temperature deviation is a characteristic value. Transmission Uncertainty, All Options Reflection Uncertainty, All Options Reflection Coefficient (linear) Reflection Coefficient (linear)

Open the catalog to page 9
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 10

Table 4. N4691B 2-Port Electronic Calibration Module Source Match Load Match 1 Temperature deviation is a characteristic value. Transmission Uncertainty, All Options Reflection Uncertainty, All Options Reflection Coefficient (linear)

Open the catalog to page 10
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 11

Table 5. N4433A4-Port Electronic Calibration Module Note: Uncertainty curves for the N4433A are created using a 2-port calibration. Multiport uncertainties are not supported at this time._ Source Match Load Match 1 Temperature deviation is a characteristic value. Transmission Uncertainty, All Options 10 H5232A 416 Full Two Port Cal Using N4433A N5232A 416 Full Two Port Cal Using N4433A Reflection Uncertainty, All Options Reflection Coefficient (linear) Reflection Coefficient (linear)

Open the catalog to page 11
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 12

Uncorrected System Performance Specifications apply to following conditions: ■ Over environmental temperature of 25 °C ±5 °C, ■ Cable loss not included in Transmission Tracking. ■ Crosstalk measurement conditions: normalized to a thru, measured with shorts on all ports, 10 Hz IF bandwidth, averaging factor of 8, alternate mode, source power set to the specified maximum power. Table 6a. Error Terms (dB), N5239A/31A/32A, All Ports, Option 200, 216 - Specifications . . . . Source Load Transmission Reflection Match Match Tracking Tracking Table 6b. Error Terms (dB), N5239A/31A/32A, All Ports,...

Open the catalog to page 12
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 13

Table 6c. Error Terms (dB), N5231A, N5232A, All Ports, Option 400, 416 - Specifications Description Directivity . Load Match ., .. „. Crosstalk Match Tracking Tracking Table 6d. Error Terms (dB), N5231A, N5232A, All Ports, Option 400,416 - Typical

Open the catalog to page 13
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 14

Test Port Output Table 7. Frequency Information, All Options Description Frequency Resolution Frequency Accuracy Assumes no variation in time. Assumes no variation in temperature. Table 8a. Maximum Leveled Power (dBm), All Models, All Ports1 Description Any port can be used as the source port. Table 8b. Maximum Leveled Power (dBm), N5231A or N5232A, All Ports1 Description Any port can be used as the source port. 14

Open the catalog to page 14
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 15

Table 9a. Power Level Accuracy (dB) at Nominal Power1, All Models, Options 200 or 216, All Ports2 Description accuracy at power other than nominal power, Power Level Accuracy (dB) at Nominal Power + Power Level Linearity (dB) 2 Any port can be used as the source port. Table 9b. Power Level Accuracy (dB) at Nominal Power1, N5231A or N5232A, Options 400 or 416, All Ports2 Description accuracy at power other than nominal power, Power Level Accuracy (dB) at Nominal Power + Power Level Linearity (dB) 2 Any port can be used as the source port.

Open the catalog to page 15
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 16

Table 10a. Power Level Linearity1 (dB), All Models, Options 200 or 216, All Ports2 - Specification 1 Referenced to nominal power. 2 Any port can be used as the source port. Table 10b. Power Level Linearity1 (dB), N5231A, N5232A Option 400,416, All Ports2 - Specification 1 Referenced to nominal power. 2 Any port can be used as the source port. Table 11a. Power Sweep Range (dB), All Models, Option 200 or 216, All Ports1 1 Any port can be used as the source port.

Open the catalog to page 16
Agilent  2-Port and 4-Port  PNA-L Network Analyzer - 17

Table 11b. Power Sweep Range (dB), N5231Aor N5232A, Option 400 or 416, All Ports1 Any port can be used as the source port.

Open the catalog to page 17