For the very latest specifications visit www.aeroflex.com
Gain
6 dB, Typical
TIME BASE (TCXO)
Temperature Stability
±1 ppm
Aging
±1 ppm per year
Accuracy
±1 ppm
Test Limit
±0.3 ppm
BATTERY
Type
Li Ion
Duration
>4 hrs continuous operation
>6 hrs, Typical
INPUT POWER (TEST SET)
Input Range
11 to 32 Vdc
Power Consumption
55 W Maximum
16 W Nominal at 18 Vdc with charged battery
Fuse Requirements
5 A, 32 Vdc, Type F
INPUT POWER (SUPPLIED EXTERNAL AC TO DC
CONVERTER)
Input Range
100 to 250 VAC, 1.5 A Max, 47 to 63 Hz
Mains Supply Voltage Fluctuations
10% of the nominal voltage
Transient Overvoltages
According to Installation Category II
ENVIRONMENTAL (TEST SET)
Use
Pollution Degree 2
Altitude
4800 meters
Operating Temperature
NOTE 2 -20°C to 55°C
Storage Temperature
NOTE 3 -30°C to 71°C
Relative Humidity
95% (±5%) from 5° to 30°C
75% (±5%) from 30° to 40°C
45% (±5%) from 40° to 55°C
ENVIRONMENTAL (SUPPLIED EXTERNAL AC TO DC
CONVERTER)
Use
Indoors
Altitude
10,000 meters
Operating Temperature
0° to 40°C
Storage Temperature
-20°C to 71°C
PHYSICAL CHARACTERISTICS
DIMENSIONS
Height
11.2 inches (28.5 cm)
Width
9.1 inches (23.1 cm)
Depth
2.7 inches (6.9 cm)
Weight (Test set only)
<8 lbs. (3.6 kg)
SUPPLEMENTAL INFORMATION
Test Set Certifications
Altitude, operating MIL-PRF-28800F Class 2
Altitude, not operating MIL-PRF-28800F Class 2
Bench Handling MIL-PRF-28800F Class 2
Blowing Dust MIL-STD-810F Method 510.4,
Procedure I
Drip-proof MIL-PRF-28800F Class 2
Explosive Atmosphere MIL-STD-810F Method 511.4,
Procedure 1
Relative Humidity MIL-PRF-28800F Class 2
Shock, Functional MIL-PRF-28800F Class 2
Vibration Limits MIL-PRF-28800F Class 2
Temp, operating NOTE 4 MIL-PRF-28800F Class 2
Temp, not operating NOTE 5 MIL-PRF-28800F Class 2
Transit Drop MIL-PRF-28800F Class 2
Safety Compliance UL-61010B-1
EN 61010-1
CSA 22.2 No 61010-1
EMC EN 61326
External AC-DC Converter Certifications
Safety Compliance UL 1950 DS
CSA 22.2 No. 234
VDE EN 60 950
EMI/RFI Compliance FCC Docket 20780 Curve “B”
EMC EN 61326
Transit Case Certifications
Drop Test FED-STD-101C Method 5007.1
Paragraph 6.3,
Procedure A,
Level A
Falling Dart Impact ATA 300 Category I
Vibration, Loose Cargo FED-STD-101C Method 5019
Vibration, Sweep ATA 300 Category I
Simulated Rainfall MIL-STD-810F Method 506.4
Procedure II of 4.1.2
FED-STD-101C Method 5009.1
Sec 6.7.1
Immersion MIL-STD-810F Method 512.4
|
|