Aeroflex 3410 Series Signal Generators - AEROFLEX - #4

/ 12


catalogue search
P. 01
P. 02
P. 03
P. 04
P. 05
P. 06
P. 07
P. 08
P. 09
P. 10
P. 11
P. 12
Pages:


See other catalogues for AEROFLEX

Text version of the page
Size The 2U rack height ensures the 3410 series occupies minimal space in a manufacturing rack or on the engineer's bench, allowing the provision of more compact test systems. The full rack width ensures easy stacking of instruments while the light weight allows for easy carrying in the laboratory or the field. SPECIFICATIONS CARRIER FREQUENCY Range 250 kHz - 2 GHz (3412) 250 kHz - 3 GHz (3413) 250 kHz - 4 GHz (3414) 250 kHz - 6 GHz (3416) Resolution 1 Hz, accuracy as frequency standard The carrier output phase can be advanced or retarded in increments of 0.036°. FREQUENCY SETTING TIME (NON-LIST MODE) After receipt of the GPIB interface deliminator (terminator), 23°C ± 5°C phase noise mode optimised >10 kHz <5.5 ms*, typ 4 ms 375 MHz, to be within 200 Hz >375 MHz, to be within 0.1 ppm phase noise mode optimised <10 kHz <3 ms*, typically 2.5 ms, 375 MHz, to be within 200 Hz <2.5 ms*, typically 2 ms, >375 MHz, to be within 0.1 ppm * For instruments fitted with option 2, add 0.5ms FREQUENCY SETTING TIME (OPTION 010 LIST MODE) After external trigger in List Mode, 23°C ± 5°C Phase Noise Mode Optimized >10 kHz <4 ms, typically 3 ms, 375 MHz, to be within 375 MHz, to be within 375 MHz, to be within -30 dBm Auto 2 GHz ±0.75 dB ± 0.50 dB 3 GHz ±1.00 dB ±0.75 dB -110 to -30 dBm >-30 dBm 6 GHz ±1.25 dB ±1.00 dB Mechanical Attenuator RF Mode -127 to -28 dBm >-28 dBm Auto 2 GHz ±0.75 dB ± 0.50 dB 3 GHz ±1.00 dB ±0.75 dB No Attenuator RF Mode >0 dBm Auto 2 GHz ±0.50 dB 3 GHz ±0.75 dB 6 GHz ±1.00 dB Level Accuracy With IQ Modulation For constant envelope modulation systems: typical standard level error ±0.15 dB For non-constant envelope modulation systems: typical standard level error ±0.25 dB Temperature Stability ±0.01 dB/°C, 3 GHz ±0.02 dB/°C, 4 GHz, ±0.02 dB/°C typical, 6 GHz RF Flatness Typical flatness at 0 dBm LEVEL SETTING TIME Electronic attenuator, Option 003 is assumed in all cases. ALC loop bandwidth ‘Moderate’ or ‘Broad’, to be within 0.3 dB Level Setting Time (Non-List Mode) After receipt of the GPIB interface deliminator (terminator), 23°C ± 5°C <4.5 ms, typically 2.5 ms Level Setting Time (Option 010 List Mode) After external trigger in List Mode, 23°C ± 5°C <3 ms, typically 1.5 ms

pageCatalog pdf di En 2012-02-07-17