Size
The 2U rack height ensures the 3410 series occupies minimal space
in a manufacturing rack or on the engineer's bench, allowing the provision
of more compact test systems. The full rack width ensures
easy stacking of instruments while the light weight allows for easy
carrying in the laboratory or the field.
SPECIFICATIONS
CARRIER FREQUENCY
Range
250 kHz - 2 GHz (3412)
250 kHz - 3 GHz (3413)
250 kHz - 4 GHz (3414)
250 kHz - 6 GHz (3416)
Resolution
1 Hz, accuracy as frequency standard
The carrier output phase can be advanced or retarded in increments of
0.036°.
FREQUENCY SETTING TIME (NON-LIST MODE)
After receipt of the GPIB interface deliminator (terminator), 23°C ± 5°C
phase noise mode optimised >10 kHz
<5.5 ms*, typ 4 ms
375 MHz, to be within 200 Hz
>375 MHz, to be within 0.1 ppm
phase noise mode optimised <10 kHz
<3 ms*, typically 2.5 ms, 375 MHz, to be within 200 Hz
<2.5 ms*, typically 2 ms, >375 MHz, to be within 0.1 ppm
* For instruments fitted with option 2, add 0.5ms
FREQUENCY SETTING TIME (OPTION 010 LIST MODE)
After external trigger in List Mode, 23°C ± 5°C
Phase Noise Mode Optimized >10 kHz
<4 ms, typically 3 ms, 375 MHz, to be within 375 MHz, to be within 375 MHz, to be within -30 dBm
Auto 2 GHz ±0.75 dB ± 0.50 dB
3 GHz ±1.00 dB ±0.75 dB
-110 to -30 dBm >-30 dBm
6 GHz ±1.25 dB ±1.00 dB
Mechanical Attenuator
RF Mode -127 to -28 dBm >-28 dBm
Auto 2 GHz ±0.75 dB ± 0.50 dB
3 GHz ±1.00 dB ±0.75 dB
No Attenuator
RF Mode >0 dBm
Auto 2 GHz ±0.50 dB
3 GHz ±0.75 dB
6 GHz ±1.00 dB
Level Accuracy With IQ Modulation
For constant envelope modulation systems: typical standard level error
±0.15 dB
For non-constant envelope modulation systems: typical standard level
error ±0.25 dB
Temperature Stability
±0.01 dB/°C, 3 GHz
±0.02 dB/°C, 4 GHz, ±0.02 dB/°C typical, 6 GHz
RF Flatness
Typical flatness at 0 dBm
LEVEL SETTING TIME
Electronic attenuator, Option 003 is assumed in all cases.
ALC loop bandwidth ‘Moderate’ or ‘Broad’, to be within 0.3 dB
Level Setting Time (Non-List Mode)
After receipt of the GPIB interface deliminator (terminator), 23°C ± 5°C
<4.5 ms, typically 2.5 ms
Level Setting Time (Option 010 List Mode)
After external trigger in List Mode, 23°C ± 5°C
<3 ms, typically 1.5 ms
|
|