For the very latest specifications visit www.aeroflex.com
ACCESSORY OPTIONS TO MAKE THE MEASUREMENTS
The user can choose from a range of accessories to suit the test
regime required.
In order for the DTF option to operate, the 2399C must be fitted with
the optional tracking generator (2399C/1) and the DTF software,
option 011.
Hardware Configurations
1. VSWR (Return Loss) only
Comprises:
RF Bridge (5 MHz to 3 GHz) 59999/170
RF cable, 0.5 m [type N (male) to type N (male)] 54351/022
Adapter type N (male) to type N (male) 54311/175
2. DTF only
Comprises:
Power Divider (type N) 54311/187
RF cable, 0.5 m [type N (male) to type N (male)] 54351/022
Calibration Load (50 Ohm) 54421/023
3. VSWR (return loss) and DTF
Comprises:
RF Bridge (5 MHz to 3 GHz) 59999/170
RF cable, 0.5 m [type N (male) to type N (male)] 54351/022
Adapter type N (male) to type N (male) 54311/175
Power Divider (type N) 54311/187
Calibration Load (50 Ohm) 54421/023
MARKER LABEL EDIT (OPTION 12)
This software option allows the user to change the marker label from
the normal numeric format to a user defined 4-digit alpha-numeric
label.
EMC ANALYZER SOFTWARE (OPTION 13)
This software option, which must be used in conjunction with
Option 04 (Quasi-peak detectors and filters) and Option 7 (Internal
pre-amplifier) which provides the user with some of the facilities
required for EMC pre-compliance testing.
Features include:
Entry of correction factors for: Test Antenna
Cable loss
Transducer characteristics
Addition of limit lines
Choice of Log or Linear frequency scales
Semi-automated operation of quasi-peak functions
Upto 3 GHz
Radiated, conducted emissions
VERSIONS, OPTIONS AND ACCESSORIES
When ordering please quote the full ordering number information.
Ordering
Numbers Versions
2399C/0 1 kHz to 3 GHz basic spectrum analyzer
2399C/1 2399C/0 with tracking generator
Options
03 High stability timebase
04 Quasi-peak detectors and filters
05 Digital resolution bandwidth filters
06 AC/DC power supply
07 Internal pre-amplifier
11 Distance to fault (DTF) software
12 Marker label edit software
13 EMC analyzer software
Supplied Accessories
Front cover
CD ROM Containing: Operation manual and
Programming manual
AC supply lead
RS-232 cable
2 x 250 V, 3.15 A fuses
Optional Accessories
Maintenance manual
AC2621 Rack mount kit
59999/170 Return loss bridge 5 MHz to 3 GHz, type N (f)
AC4250 50 Ù N to 75 Ù BNC
AC5008 DC block N type
AC8700 UHF antenna
80010 Soft carry case
|
|