Number of Steps: Selectable from 1 to 10,000 steps
Step Size: Selectable, any size consistent with resolution of unit
Stop Sweep: Causes internal sweep to halt immediately. Return control
to command level.
MODULATION
FM
Frequency Peak Deviation 3 dB
Range GHz MHz Bandwidth
0.01 - 3.999 0 to 1 10 Hz to 300 kHz
1.15 - 3.999 0 to 10 10 Hz to 300 kHz
AM
Frequency Depth 3 dB
Range MHz % Bandwidth
10 - 180 0 to 90 10 Hz to 50 kHz
REMOTE PROGRAMMING CONTROL INTERFACE
44 Bits Parallel BCD TTL Compatible; Positive True with Strobe. Mating
Connector: 3M P/n 3564-1000. In addition to standard Interface
IEEE-488-1978, all functions controlled from the front panel, with the
exception of the power line switch, are programmable with the same
accuracy and resolution as in manual mode.
GENERAL
Operating temperature range
0° to 50°C
Power Requirements
120/250 VAC 48 to 440 Hz, 250 Watts
Weight
50 lbs. (22.7 kg)
Dimensions
16.75" W x 5.22" H x 23.88" D (42.55 x 13.26 x 60.66 cm)
OPTIONS
OPTION 101, ADDED DDS FOR ENHANCED FREQUENCY RESOLUTION,
MODULE A5
Frequency Opt 101
Range MHz Hertz
10 MHz to 2.3 GHz 0.1
2.3 to 4 GHz 0.1
4.6 to 9.2 GHz 0.2
9.2 to 18.4 GHz 0.4
The option 101 limits the switching speed to 1.2 ěsec.
OPTION 120, FM MODULATION
Frequency Peak Deviation +/- MHz
Range MHz Wide Narrow
50-69 1.5 0.15
70-139 0.75 0.075
140-279 1.5 0.15
280-559 3 0.3
560-1149 6 0.6
1150-2299 12 1.2
2300-4599 24 2.4
4600-9199 48 4.8
9200-18399 96 9.6
OPTION 121, PROGRAMMABLE FM, MAINTAINS CONSTANT DEVIATION
ACROSS FREQUENCY RANGE
Frequency Peak Deviation
Range MHz MHz
50-1149 0.01, 0.1, 1
1150-18399 0.1, 1, 10
External only FM 3 dB
coupling mode Bandwidth
DC DC to 5 MHz
AC 50 Hz to 5 MHz
OPTION 122, PULSE MODULATION
On/OFF ratio
60 dB
Rise/Fall time
10 nsec
OPTION 125, FAST ATTENUATOR, SOLID STATE
Frequency Range
0.5 to 18 GHz
Attenuation Range
0 to 60 dB
Attenuation Increment
0.25 dB
Switching Time
1 ěsec max
OPTION 128, SLOW ATTENUATOR, MECHANICAL
Frequency Range
10 MHz to 18.4 GHz
Attenuation Range
0 to 120 dB
Attenuation Increment
1 dB
Switching Time
20 msec max
Contact the factory for non-standard options such as phase modulation