You may also be interested in
Testing machine, Noise analyzer, Protocol analyzer, Antenna, Chassis
Text version of the page
ON/OFF ratio
Greater than 80 dB.
Envelope delay
< 10 ės, typically 6 ės.
Rise/fall time
Less than 13 ės to 70 dBc.
IF OUTPUT
An IF output is available on the rear panel which is modulated by the
selected digital or vector modulation. The IF output can be inhibited by
software control. The IF output can be used to provide modulated carriers
at higher frequencies by external frequency conversion. The RF
output from the front panel connector can be used as an LO for external
frequency conversion.
MODULATION OSCILLATOR
Frequency range
0.1 Hz to 500 kHz.
Selection
By keyboard entry of data. Variation by keys and by rotary control.
Indication
7 digits with annunciators.
Resolution
0.1 Hz.
Frequency accuracy
As frequency standard.
Distortion
Less than 0.1% THD in sine wave mode at frequencies up to 20 kHz.
Alternative waveform
A triangular wave is available in addition to the sine wave for frequencies
up to 100 kHz.
Signaling tones
The modulation oscillator can be used to generate sequential (up to
16 tones) or sub-audible signaling tones in accordance with EIA, ZVEI,
DZVEI, CCIR, EURO 1, EEA, NATAL and DTMF* standards.
Facilities are also available for creating and storing user defined tone
systems.
* Requires second modulation oscillator (option 001) to be fitted.
EXTERNAL MODULATION
Two independent inputs on the front panel with BNC connectors, EXT
MOD 1 and EXT MOD 2. The modulation is calibrated with 1.414 V
peak (1 V RMS sine wave) applied. Input impedance 100 kŲ nominal.
MODULATION ALC
The EXT MOD 1 and EXT MOD 2 modulation inputs can be levelled by
an ALC system.
Level Range
1 V to 2 V peak (0.7 to 1.4 V RMS sine wave).
Distortion
Less than 0.1% additional distortion for frequencies up to 20 kHz (typically
less than 0.1% up to 50 kHz).
1 dB Bandwidth
Typically 10 Hz to 500 kHz.
LF OUTPUT
Front panel BNC connector. The output may be configured in the LF
Generator Mode to give an output from the internal modulation oscillator
and in the LF Monitor Mode to give an output from the internal
modulation signal paths.
Selection
By keyboard entry of data.
Variation by keys and by rotary control.
Indication
7 digits with unit annunciators for frequency and 4 digits with unit
annunciators for level.
Level
100 ėV to 5 V RMS with a load impedance of greater than 600 Ų.
100 ėV to 1.4 V RMS with a load impedance of greater than 50 Ų.
Source impedance
5.6 Ų nominal.
Level accuracy at 1 kHz
With a load impedance of greater than 10 kW: LF ą5% for levels
above 50 mV LF ą10% for levels from 500 ėV to 50 mV.
Frequency Response
Typically <ą1 dB from 0.1 Hz to 300 kHz.
SWEEP
Control modes
Start/stop values of selected parameter. Number of steps. Time per
step.
Step time
1 ms to 20 s per step.
Sweep ramp
Synchronized analog ramp with a nominal amplitude of 0 to 10 V peak
on rear panel BNC connector.
Markers
User selectable markers for frequency or level provide an indication
when specified parameter values have been reached. Output 0 V to
+5 V from 600 Ų on rear panel BNC socket.
Trigger
Rear panel BNC connector. Applying 0 V or a switch closure starts the
sweep. Connector is internally connected via 10 kŲ pull-up resistor to
+5 V.
FREQUENCY STANDARD (OCXO)
Frequency
10 MHz.
Temperature stability
Better than ą5 in 108 in the operating range of 0 to 50°C.
Warm up time
Within 2 in 107 final frequency within 10 minutes from switch on at
20°C ambient.
Ageing rate
Better than 2 in 107 per year.
Output
Rear panel BNC socket provides an output at frequencies of 1, 5 or
10 MHz with a nominal 2 V pk-pk level into 50 Ų. Output can be dis
|
|