eXpert 760x Specifications
Model
7601
7602
7603
7605
Table Top
Table Top
Table Top
Table Top
Load Capacity
lbf
225
500
1,000
2,250
kN
1
2.5
5
10
kgf
100
250
500
1,125
Maximum Speed
in/min
40
20
40
20
mm/min
1,016
508
1,016
508
Minimum Speed
in/min
0.0005
0.0005
0.0001
0.0001
mm/min
0.012
0.012
0.0025
0.0025
Maximum Force
lbf
175
450
1,000
2,250
at Full Speed
kN
0.75
2
5
10
Position Control
min
4.2
2.1
1.1
0.68
Resolution
jm
0.107
0.053
0.028
0.017
Total Crosshead
in
30
30
34
34
Travel1
mm
762
762
864
864
Total Vertical
in
34
34
37.5
37.5
Test Space2
mm
864
864
952
952
Lateral Test
in
3.4
3.4
3.4
3.4
Space
mm
87
87
87
87
Space Between
in
__
--
--
--
Columns
mm
--
--
--
Height
in
40
40
49
49
mm
1,016
1,016
1,245
1,245
Width
in
16
16
17
17
mm
406
406
432
432
Depth
in
12
12
19.8
19.8
mm
305
305
503
503
Weight
lbf
45
55
200
225
kgf
20.5
25
90
102
Maximum Power
VA
100
100
300
300
Single Phase
VAC
100-240
100-240
110,220
110,220
Voltage
Hz
50,60
50,60
50,60
50,60
Load Measurement Accuracy: +/- 0.5% of reading down to 1/100 of load cell capacity. Meets or exceeds ASTM E4, BSENISO7500-1 : 2004, DIN 51221 and JIS B7721standards. ADMET self-identifying load cells are available.
Strain Measurement Accuracy: +/- 0.5% of reading down to 1/50 of full scale with ASTM E83 class B extensometers. Meets or exceeds ASTM E83 and BSENISO9513 : 2002 standards.
Notes:
1) Total Crosshead Travel is calculated without load cells, grips and fixtures. Longer strokes can be accommodated by ordering an extended column frame.
2) Total Vertical Test Space is the distance from the top surface of the base platen to the bottom surface of the moving crosshead, excluding load cell, grips and fixtures. Larger openings can be accommodated by ordering an extended column frame.
Copyright 2006. All Rights Reserved. ADMET, Inc. Norwood, MA 02062

pageCatalog pdf di En 2012-05-21-25