| | | eXpert 560x Specifications | | |
| | | |
| | | | | | | | | | | | | | | | | Model | 5601 | 5601QP | 5602 | 5603 | 5606 | 5607 | Two Column Frame Option | | | | Load Capacity | lbf | 225 | 100 | 500 | 1,000 | 1,400 | 1 | | | | | | kN | 1 | 0.5 | 2.2 | 4.5 | 6.2 | 0.004 | Any actuator | | | | | kgf | 100 | 50 | 225 | 450 | 630 | 0.45 | can be mounted | | | | Maximum Speed | in/min | 25 | 180 | 7 | 20 | 0.9 | 300 | -- | | | | | mm/min | 635 | 4570 | 178 | 504 | 23 | 7,620 | | | | | Minimum Speed | in/min | 0.001 | 0.005 | 0.0005 | 0.001 | 0.00001 | 0.01 | -- | | | | | mm/min | 0.025 | 0.05 | 0.012 | 0.025 | 0.00025 | 0.25 | -- | | | | Maximum Force | lbf | 225 | 20 | 500 | 800 | 1,400 | 1 | -- | | | | at Full Speed | kN | 1 | 0.1 | 2.2 | 3.6 | 6.2 | 0.004 | -- | | | | Position Control | u in | 2.9 | 14.6 | 0.87 | 2.2 | 0.1 | 35 | -- | | | | Resolution | um | 0.073 | 0.38 | 0.022 | 0.057 | 0.0025 | 0.88 | -- | | | | Total Crosshead | in | 12 | 12 | 12 | 12 | 12 | 125 | -- | | | | Travel | mm | 305 | 305 | 305 | 305 | 305 | 305 | -- | | | | Total Vertical | in | 31 | 31 | 31 | 31 | 31 | 31 | 32 | | | | Test Space1 | mm | 787 | 787 | 787 | 787 | 787 | 787 | 813 | | | | Lateral Test2 | in | 3.5 | 3.5 | 3.5 | 3.5 | 3.5 | 3.5 | -- | | | | Space | mm | 89 | 89 | 89 | 89 | 89 | 89 | -- | | | | Space Between | in | „ | -- | „ | -- | -- | -- | 16.5 | | | | Columns | mm | | -- | -- | -- | -- | -- | 419 | | | | Height3 (Vertical Frame) | in | 53.3 | 53.3 | 53.3 | 53.3 | 53.3 | 53.3 | 54.3 | | | | | mm | 1354 | 1354 | 1354 | 1354 | 1354 | 1354 | 1379 | | | | Width (Vertical Frame) | in | 18.9 | 18.9 | 18.9 | 18.9 | 18.9 | 18.9 | 19.5 | | | | | mm | 480 | 480 | 480 | 480 | 480 | 480 | 495 | | | | Depth (Vertical Frame) | in | 15.1 | 15.1 | 15.1 | 15.1 | 15.1 | 15.1 | 16 | | | | | mm | 384 | 384 | 384 | 384 | 384 | 384 | 406 | | | | Weight | lbf | 45 | 45 | 48 | 70 | 70 | 45 | 80 | | | | | kgf | 20.5 | 20.5 | 22 | 32 | 32 | 20.5 | 36 | | | | Maximum Power | VA | 100 | 350 | 100 | 250 | 100 | 100 | -- | | | | Single Phase | VAC | 100-240 | 100-240 | 100-240 | 110,220 | 100-240 | 100-240 | -- | | | | Voltage | Hz | 47-63 | 47-63 | 47-63 | 50,60 | 47-63 | 47-63 | -- | | | | | | | | | | | | | | | | |
| | | |
| | | Load Measurement Accuracy: +/- 0.5% of reading down to 1/100 of load cell capacity. Meets or exceeds ASTM E4, BSENISO7500-1 : 2004, DIN 51221 and JIS B7721standards. ADMET self-identifying load cells are offered with all systems. | | |
| | | |
| | | Strain Measurement Accuracy: +/- 0.5% of reading down to 1/50 of full scale with ASTM E83 class B extensometers. Meets or exceeds ASTM E83 and BSENISO9513 : 2002 standards. | | |
| | | |
| | | Notes: 1) Total Vertical Test Space is the distance from the top surface of the base platen to the bottom surface of the moving crosshead, excluding load cell, grips and fixtures when the actuator is adjusted to its highest point on the vertical column. Larger openings can be accommodated by ordering an extended column. 2) Larger Lateral Test Spaces can be special ordered. 3) Height is measured when the actuator is adjusted to its highest point on the vertical column. 4) Noise level measured at full speed and 1m from machine is 60db. 5) Tension only applications. Maximum stroke in compression is 3 in. | | |
| | | |
| | | Copyright (c) 2005-2008. ADMET, Inc. Norwood, MA, USA. All Rights Reserved. | | |
| | | |
| | | | | | | | | | | | |