eXpert 560x Universal Testing Machines - ADMET - #4

/ 4


catalogue search
P. 01
P. 02
P. 03
P. 04
Pages:


See other catalogues for ADMET

Text version of the page
eXpert 560x Specifications
Model
5601
5601QP
5602
5603
5606
5607
Two Column Frame Option
Load Capacity
lbf
225
100
500
1,000
1,400
1
kN
1
0.5
2.2
4.5
6.2
0.004
Any actuator
kgf
100
50
225
450
630
0.45
can be mounted
Maximum Speed
in/min
25
180
7
20
0.9
300
--
mm/min
635
4570
178
504
23
7,620
Minimum Speed
in/min
0.001
0.005
0.0005
0.001
0.00001
0.01
--
mm/min
0.025
0.05
0.012
0.025
0.00025
0.25
--
Maximum Force
lbf
225
20
500
800
1,400
1
--
at Full Speed
kN
1
0.1
2.2
3.6
6.2
0.004
--
Position Control
u in
2.9
14.6
0.87
2.2
0.1
35
--
Resolution
um
0.073
0.38
0.022
0.057
0.0025
0.88
--
Total Crosshead
in
12
12
12
12
12
125
--
Travel
mm
305
305
305
305
305
305
--
Total Vertical
in
31
31
31
31
31
31
32
Test Space1
mm
787
787
787
787
787
787
813
Lateral Test2
in
3.5
3.5
3.5
3.5
3.5
3.5
--
Space
mm
89
89
89
89
89
89
--
Space Between
in
--
--
--
--
16.5
Columns
mm
--
--
--
--
--
419
Height3 (Vertical Frame)
in
53.3
53.3
53.3
53.3
53.3
53.3
54.3
mm
1354
1354
1354
1354
1354
1354
1379
Width (Vertical Frame)
in
18.9
18.9
18.9
18.9
18.9
18.9
19.5
mm
480
480
480
480
480
480
495
Depth (Vertical Frame)
in
15.1
15.1
15.1
15.1
15.1
15.1
16
mm
384
384
384
384
384
384
406
Weight
lbf
45
45
48
70
70
45
80
kgf
20.5
20.5
22
32
32
20.5
36
Maximum Power
VA
100
350
100
250
100
100
--
Single Phase
VAC
100-240
100-240
100-240
110,220
100-240
100-240
--
Voltage
Hz
47-63
47-63
47-63
50,60
47-63
47-63
--
Load Measurement Accuracy: +/- 0.5% of reading down to 1/100 of load cell capacity. Meets or exceeds ASTM E4, BSENISO7500-1 : 2004, DIN 51221 and JIS B7721standards. ADMET self-identifying load cells are offered with all systems.
Strain Measurement Accuracy: +/- 0.5% of reading down to 1/50 of full scale with ASTM E83 class B extensometers. Meets or exceeds ASTM E83 and BSENISO9513 : 2002 standards.
Notes:
1) Total Vertical Test Space is the distance from the top surface of the base platen to the bottom surface of the moving crosshead, excluding load cell, grips and fixtures when the actuator is adjusted to its highest point on the vertical column. Larger openings can be accommodated by ordering an extended column.
2) Larger Lateral Test Spaces can be special ordered.
3) Height is measured when the actuator is adjusted to its highest point on the vertical column.
4) Noise level measured at full speed and 1m from machine is 60db.
5) Tension only applications. Maximum stroke in compression is 3 in.
Copyright (c) 2005-2008. ADMET, Inc. Norwood, MA, USA. All Rights Reserved.